Temperature scale: Celsius or Fahrenheit user-selectable
Resolution: 1°C or 1°F (or 0.1°C or 0.1°F) user-selectable
Inputs: K-type thermocouples
Measurement Range: -50°C to 1300°C, (-58°F to 2000°F)
Meter Operating Range: For use in 0°C to 50°C environment (32°F to 122°F)
Accuracy: (18°C to 28°C ambient)
±(0.3% rdg + 1°C) -50°C to 1000°C
±(0.5% rdg + 1°C) 1000°C to 1300°C
±(0.3% rdg + 2°F) -58°F to 2000°F
The accuracy specification does not include type K thermocouple probe accuracy.
Temperature Coefficient: 0.1 times the applicable accuracy specification per °C from 0°C to 18°C and 28°C to 50°C.
Error Offset: Errors contributed by one particular thermocouple may be practically eliminated, over a limited measurement range, using a recessed front panel OFF-SET adjustment on the meter and an accurate temperature standard when making the adjustment. Although making that adjustment may increase measurement errors made using other thermocouples.
Maximum Input: 60V dc or 24V rms ac between the input pins or any input and ground.
Display: 3½ digit liquid crystal display (LCD) with maximum reading of 1999
Sampling Rate: 3 times per second
Battery: One 9V battery (NEDA 1604)
Battery Life: 200 hours typical with carbon zinc battery
Dimensions: 147(H) x 70(W) x 39(D) mm
Weight: 260g