10, 100, and 250 MHz models
Division factor variable from 2 to 255, 2 to 999, or 2 to 65535
Variable output pulse widths
Low jitter
TTL and ECL levels available
BNC input and output connectors
Input of 120/240 Volts, 50-60 Hz prime power
The OE-FD series of digital frequency dividers will divide the pulse
repetition frequency of an input pulse train by a factor (N) which is variable from 2 to 999
for the 10 MHz Model OE-FD1.PS, 2 to 255 for the 100 MHz Model OE-FD2.PS, and
2 to 65535 for the 250 MHz Model OE-FD3.PS.
|
Model |
OE - FD1.PS |
OE - FD2.PS |
OE - FD3.PS |
Input frequency: |
0 - 10 MHz |
0 - 100 MHz |
0 - 250 MHz |
Division factor (N): |
2 to 999 |
2 to 255 |
2 to 65535 |
Input level: |
TTL (0 and 3 - 5V) |
ECL (-0.8 and -1.6V) |
Input termination: |
50 Ohm or 1 kOhm to ground, switchable |
50 Ohm to -2V |
Input pulse width: |
> or = 10 ns |
> or = 1 ns |
Output level: |
TTL (0 and 3 - 5V) |
ECL (-0.8 and -1.6V) |
Outputs: |
Main Output: fOUT = fIN/N |
OUT1: fOUT1 = fIN/N
OUT2: fOUT2 = fOUT1/2 |
Output pulse width: |
50 ns to 50 ms |
5 ns to 5.0 ms |
OUT1: one input period, i.e. PWOUT1 = 1/fIN
OUT2: 50% duty cycle, i.e. PWOUT2 = 1/fOUT1 |
Jitter: |
< or = 100 ps |
Connectors: |
BNC |
Prime power: |
120/240 Volts, 50 - 60 Hz |
Dimensions (H x W x D): |
102 mm x 203 mm x 305 mm
(4" x 8" x 12") |
Temperature range: |
+ 10° to + 40° C |
Options:
ECL option: Input and output on OE-FD1.PS and OE-FD2.PS operate at ECL levels
(rather than at standard TTL-levels)
DECL option: Inputs and outputs are differential ECL
EP option: Provides complementary output pulses.
Our application engineers and technical staff will
be happy to help you with any further enquiries. Monday through
Friday, 9 am to 6 pm, Singapore Standard Time, call (65)-4409746 or email us