The following list details the status of the medical equipment that will be affected by the millennium bug, as it were. Note that included are equipment that will be affected functionally or non-functionally. So far, there are 37 models at issue, giving some 250 pieces of equipment. This list will be kept being updated as more information comes to light.
|
|
|
|
|
|
|
|
|
|
|
To contact y2k team at Toshiba, click here |
|
|
Sorting by Expected Delivery Date (EDD) function returns results in the following order ’00, ’97, ’98, ’99. |
OR A Year 2000 upgrade kit is available for later model units (dependent upon serial number). Solution will become available in 1st Qtr 1999. |
|
|
|
To contact y2k team at Toshiba, click here |
|
|
Sorting by Expected Delivery Date (EDD) function returns results in the following order ’00, ’97, ’98, ’99. |
OR A Year 2000 upgrade kit is available for later model units (dependent upon serial number). Solution will become available in 1st Qtr 1999. |
|
|
|
e-mail of Y2K team: y2k@radiometeramerica.com |
(Al-Darhaly) |
|
Also, all power off tests failed (Y2000, Y2001) A certificate of compliance is available. |
|
|
|
|
If your computer is still unable to achieve Year 2000 compliance, please contact LUNAR's Customer Support Department at 1-800-334-5831, (608) 274-8210 [FAX] Lunar's Y2K e-mail |
|
|
|
|
|
|
|
|
|
|
This system will be affected by the Year 2000 change. |
|
|
|
|
Status of Spacelabs' Medical Equipment |
(Al- Muttahida) |
|
This equipment is not y2k compliant. Requires Software. Monitors with Tabular Trends function may momentarily display an invalid number for the date day at midnight on the last day of each month. The display will automatically update and correct the date within 1 minute. |
|
|
(Mainframe Monitor) |
|
Status of Spacelabs' Medical Equipment |
(Al- Muttahida) |
|
This equipment is not y2k compliant. Requires Software. Monitors with Tabular Trends function may momentarily display an invalid number for the date day at midnight on the last day of each month. The display will automatically update and correct the date within 1 minute. |
|
|
|
|
Y2K Page |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
List of Affected Computed Tomography (CT) Systems To contact y2k team at Toshiba, click here |
|
|
Operational problem. A date after
12/31/99 cannot be entered.
The equipment will not boot after 12/31/99. An examination reservation for a date after 12/31/99 cannot be made. |
A Year 2000 upgrade kit will be
available by 1st quarter 1999.
|
|
|
OGP-7101 |
Status of Monitors |
|
= 7 |
|
|
|
|
|
List of Ultrasound Instruments Contact Shimadzu If you would like additional information on Shimadzu's Year 2000 compliance, fax 81-3-3219-5710, or send an e-mail to imd_sci@ group.shimadzu.co.jp |
Al-Tibbiya |
|
|
|
|
photometer |
|
Contact Shimadzu If you would like additional information on Shimadzu's Year 2000 compliance, fax 81-3-3219-5710, or send an e-mail to imd_sci@ group.shimadzu.co.jp |
Al-Tibbiya |
|
|
|
|
|
(Classified by Company as a Historical Product) |
Dynamic Imaging Y2K Product List
Andrew Johnston, Quality Systems Manager
|
|
|
|
For systems more than 7 years old contact Dynamic Imaging for a quotation. |
|
|
(Classified by Company as a Historical Product) |
Dynamic Imaging Y2K Product List
Andrew Johnston, Quality Systems Manager
|
|
|
|
For systems more than 7 years old contact Dynamic Imaging for a quotation. |
|
|
(Classified by Company as a Historical Product) |
Dynamic Imaging Y2K Product List
Andrew Johnston, Quality Systems Manager
|
|
|
|
For systems more than 7 years old contact Dynamic Imaging for a quotation. |
|
|
|
Nellcor Puritan Bennet Y2K Site (including a search facility) |
|
|
|
|
|
|
|
|
|
|
|
|
|
(Repeat) |
|
(The company has been acquired by Picker) |
|
|
1. Impossible to enter patient date of birth.
2. Garbage patient age. |
Installation at no charge for Full Service Contract Customers. Solution in V3.2 Important note: installation of the solution prior to 1 January 2000 cannot be guaranteed for those customers who contact Elscint after 1 July 1999. |
|
|
|
|
|
|
|
Obsolete, will not be tested for Y2K compliance.
These products will no longer be supported by AVL as of Jan 1, 2000. |
|
|
|
Hitachi Medical Corporation of America, Hitachi Medical Systems Y2K Site Contact Hitachi Medical Systems Hitachi Medical Corporation of America Hitachi Medical Systems 660 White Plains Road Tarrytown, NY 10591-5107 Phone: (914) 524-9711 or (800) 852-2080 Fax: (914) 524-9716 Email: info@ hitachiultrasound.com |
|
|
No formal corrective measure is available, however, in
order to alert the operator of this potential condition, HMCA will issue
a caution label to be supplied at the customer’s request to HMCA. TO
RECEIVE A CAUTION LABEL:
|
|
|
|
|
Sysmex Y2K Site For more detailed information, please call the Technical Support Center at: 1-888-879-7639 or e-mail |
|
|
|
|
|
|
|
|
|
|
Test Results: Y2000 Power on = Pass, Y2000 Power off = Pass, Y2001 Power on = Pass, Y2000 Power off = Pass, 30/10/2000 Monday = Not Applicable, Leap Year OK? = Fail, 9/9/99 OK? = Pass. Leap year OK with power off but not power on. |
|
|
|
|
Beckman Coulter Y2K Site |
|
Product is Year 2000 compliant for the intended use and primary functionality: minor functionality subject to operational notes. Operational Notes: Operational Notes P/N 4276750 will be available January 1999. Hardware workaround required for AMI/DTX SCAT386 and AMI/ARROW HAUPPAUGE 386(computer off, date will reset to 1980). Inferencing Rule: Two digit years with a value equal to or greater than 80 imply 19XX; those with a value less than 38 imply 20XX. Years 38 thru 79 not recognized. |
|
|
|
|
|
|
|
|
||
|
|
|
Fritz-Berne-Strase 40 - 81241 Munchen Tel: ++49 - (0)89 - 83 94 09 - 0 Fax: ++49 - (0)89 - 83 94 09 - 43 e-mail: Lohmeier@ Lohmeier.com |
|
|
|
|
|
Monitor |
|
Fritz-Berne-Strase 40 - 81241 Munchen Tel: ++49 - (0)89 - 83 94 09 - 0 Fax: ++49 - (0)89 - 83 94 09 - 43 e-mail: Lohmeier@ Lohmeier.com |
|
|
|
|
|
|
|
Contact Shimadzu If you would like additional information on Shimadzu's Year 2000 compliance, fax 81-3-3219-5710, or send an e-mail to imd_sci@ group.shimadzu.co.jp |
Al-Tibbiya |
|
Print Certificate |
|
|
phoresis System |
Catalogue No. 1260/
|
Rita Ellerbrook
|
|
There are two parts to the question of Year 2000 Compliance,
software and hardware. The software is written by Helena Laboratories.
All software upgrades are available from us at no cost. The latest version
of CliniScan 2 software is Rev. H. Older versions of the software have
not been verified as being able to handle Year 2000, and an upgrade to
Rev. H is recommended.
The CliniScan 2 software has been found to be compliant for the year 2000 with one exception. Stored data can not be retrieved by date in one pass across the millennium. You must retrieve data from before 01/01/2000 in one pass, and data after 12/31/1999 in a second pass. Data can be retrieved across the millennium by patient ID or other parameter, with no regard for date as always. Instructions for data retrieval by date over a range including 1/1/2000 are found at the end of this letter. A summary of the activities that this software performs correctly is listed below.
NOTE: Because of the possibility of the date improperly rolling over, do not create data at the time of the date change. Setting the Date:
Retrieving Data by Date Range:
|
|
Technicon DAX |
Richard Kaplan Phone: 914-631-8000 |
Medix |
|
S/W Version 4.03, 4.10 Date Format: mm/dd/yy Technicon DAX (UIW): Compliant software projected to be
released by 4/1/1999.
|
|
|
|
|
|
(Acoustic Imaging) |
|
Testing/Results
The testing methodology encompassed full system functionality testing both during and after the year 2000 transition. The testing investigated several issues; the rollover and retention of the time and date from 12/31/99 to 1/1/00, the proper identification and use of leap years subsequent to 1/1/00, i.e. 2000, 2004 etc., the real time functionality of the unit including display information and transducer operation, the data archiving functionality of the system, the two dimensional and velocity measurement functionality of the system, and the calculation package functionality of the system Completion of the aforementioned testing has indicated
that subject systems will roll over to January 1, 2000 but will not retain
this information after the system is powered off. The testing did indicate
however, that once the date and time had been reset to January 1, 2000
or later, then the system would perform as expected including the functionality
as indicated above.
|
Corrective Action
Each customer must install the enclosed software upgrade version 2.18 to assure the latest and tested version of the software is installed. To install the software, please follow these instructions: Turn the system off. |
|
33 |
|
If it is Version 15, then comments herein are valid. |
|
|
|
|
|
|
|
3002 IH |
|
|
|
|
|
|
|
|
Bridgend Road Industrial Estate Dingwall Ross-shire IVV15 9QF Scotland Tel.: +44 1349 862388 Fax: +44 1349 864146 e-mail: valeriel@albyn.com |
|
|
|
|
|
|
|
|
Tel.: 5664008 Fax: 5665093 |
|
1. The date after 1 January 2000 is not correct. 2. The birthdate is not correct crossing the millennium. 3. The LMP date is not correct crossing the millennium. |
1. The operator is able to correct the date permanently. 2. The operator has to manually compute the birthdate between 1 January 2000 and 30 September 2000. 3. The operator has to manually compute the LMP date between 1 January 2000 and 30 September 2000. |
|
|
|
Y2K Specific Contact Information 3135 Quarry Road Telford, PA 18969 USA Tel.: (215) 721 5400 Fax: (215) 723 5934 e-mail: brennanj@nad.com |
Test results 1. (stand alone) pass 2. (data in) N/A 3. (data out) pass Note: A third party device may request a time and date message which contains a two digit year. N.A.D., Inc. cannot guarantee compliance with other manufacturer's products that are attached to or used with their anaesthesia machines. The Operating Room Data Manager (ORDM) which is commonly attached to a Narkomed Anesthesia machine and utilized with off-line software is not Year 2000 compliant. The ORDM will represent the first two digits of the patient's birth date as '19'. If your software is earlier than that listed in Table 1 we recommend an upgrade of the software and hardware, if needed, to achieve a level of compliance with our published testing protocol. In the U.S. contact DrägerService at (800) 543-5047. Narkomed Anesthesia machines do not utilize date or time information in their life support or drug delivery functions. |
|||