Medical systems that will be affected
by Y2K problem and their numbers

  The following list details the status of the medical equipment that will be affected by the millennium bug, as it were. Note that included are equipment that will be affected functionally or non-functionally. So far, there are 37 models at issue, giving some 250 pieces of equipment. This list will be kept being updated as more information comes to light.

 

Amjad Jaimoukha
Amman, 22nd October 1999


No.
System
Model No.
Manufacturer
Local Agent
No. in Service
Potential Effects and Status
Solution and Comments
1
Ultrasonic Scanner
SSA-250A
ToshibaY2K Page
To contact y2k team at Toshiba, click here
Medical Systems
2
When the first day of the Last Menstrual Period (LMP) is in 1999 and the Expected Delivery Date (EDD) is in 2000, the system cannot calculate the Diagnostic Gestational Age (DGA).  DGA will be displayed as "   W   D".
Sorting by Expected Delivery Date (EDD) function returns results in the following order ’00, ’97, ’98, ’99.
Functional workaround - operational instructions.
OR
A Year 2000 upgrade kit is available for later model units (dependent upon serial number).
Solution will become available in 1st Qtr 1999.
2
Ultrasonic Scanner
SSA-250B
ToshibaY2K Page
To contact y2k team at Toshiba, click here
Medical Systems
8
When the first day of the Last Menstrual Period (LMP) is in 1999 and the Expected Delivery Date (EDD) is in 2000, the system cannot calculate the Diagnostic Gestational Age (DGA).  DGA will be displayed as "   W   D".
Sorting by Expected Delivery Date (EDD) function returns results in the following order ’00, ’97, ’98, ’99.
Functional workaround - operational instructions.
OR
A Year 2000 upgrade kit is available for later model units (dependent upon serial number).
Solution will become available in 1st Qtr 1999.
3
Blood  Gas Analyzer
ABL300
Radiometer America Y2K Site
e-mail of Y2K team:
y2k@radiometeramerica.com
Al-Ittihad
(Al-Darhaly)
6
leap years are not recognized by ABL300
Also, all power off tests failed (Y2000, Y2001)
A certificate of compliance
is available.
The date must be adjusted on 29/2/2000.
4
Bone Densitometer
DPXplus
Lunar Corporation Y2K Site
If your computer is still unable to achieve Year 2000 compliance, please contact LUNAR's Customer Support Department at 1-800-334-5831, (608) 274-8210 [FAX]
Lunar's Y2K e-mail
National
1
LUNAR systems installed prior to December 1997 may not have the latest software version and thus not be Year 2000 compliant. The system at the Ministry was installed on 10/2/1998, according to the official forms of the manufacturer. Once written confirmation is obtained, the status of this system will be upgraded to Y2K compliant.
You will need to update your system to become year 2000 compliant.
5
Gamma Camera
MicroDELTA
Siemens' Y2K Site, Product Information, Nuclear Section
Wa'il Abd Al-Hadi
1
SW-version  required V7.2 / V3.2
This system will be affected by the Year 2000 change.
Operator will be required to use "Study Utilities" to modify the date field daily. Operator will be required to enter age manually within protocols that require age. Please contact Year 2000 representative for further information. He will be able to recommend you the solution best suited for your configurati
6
PC2 SYSTEM CENTRAL DISPLAY (Mainframe monitor)
90315
Spacelabs
Status of Spacelabs' Medical Equipment
The United Company
(Al-
Muttahida)
2
Software version 3.66.92E
This equipment is not y2k compliant. Requires Software. Monitors with Tabular Trends function may momentarily display an invalid number for the date day at midnight on the last day of each month. The display will automatically update and correct the date within 1 minute.
S/W update is available. V3.67.04EN required for compliance. Solution software will be available by January 1999. It can be installed by the customer.
7
PC1 Bedside Monitor
(Mainframe Monitor)
90303B
Spacelabs
Status of Spacelabs' Medical Equipment
The United Company
(Al-
Muttahida)
10
Software version 3.51.46EN
This equipment is not y2k compliant. Requires Software. Monitors with Tabular Trends function may momentarily display an invalid number for the date day at midnight on the last day of each month. The display will automatically update and correct the date within 1 minute.
S/W update is available. V3.67.04EN required for compliance. Solution software will be available by January 1999. It can be installed by the customer.
8
Bedside Monitor
OMP-7201/K
Nihon Kohden
Y2K Page
Al-Ittihad Company
42
This equipment does not support dates after December 31, 1999. An incorrect year will be displayed but the instrument will function normally in other respects.
The manufacturer cannot provide a solution because the software ROMs can no longer be procured.
9
MRI 
Prestige 2T
Elscint Y2K Site (The Company has been acquired by Picker)
Philadelphia
1
Needs a solution package.
Contact manufacturer.
10
Ultrasound
-
HP
Al-Lawazim al-'ilmiya
3
Have to be reset by operator on first use after 2000.
 
11
Computed Tomography - Scan (CT-Scan)
Xvision/GX
Toshiba Y2K Page
List of Affected Computed Tomography (CT) Systems
To contact y2k team at Toshiba, click here
Medical Systems
2
Operational problem. A date after 12/31/99 cannot be entered. 
The equipment will not boot after 12/31/99. 
An examination reservation for a date after 12/31/99 cannot be made. 
Dates in 20XX will be displayed as 19100 on the "EL" panel.
A Year 2000 upgrade kit will be available by 1st quarter 1999.
Once the kit is installed all dates must be entered using 4 digits. Dates entered with 2 digits will be interpreted as 19XX.
12
Central Monitor
OGP-7201
OGP-7101
Nihon Kohden Y2K Page
Status of Monitors
Al-Ittihad Company
OGP-7101
= 7
If the backup battery is charged and functioning properly, the instrument is able to recognize and calculate dates after 1999 as accurately as dates prior to year 2000. If the backup battery is discharged or  not funtioning properly, the instrument must be reset manually to year 2000 and is then year 2000 compliant.
 
13
Ultrasound System
SDL-310B 
Shimadzu Y2K Site
List of Ultrasound Instruments
Contact Shimadzu
If you would like additional information on Shimadzu's Year 2000 compliance, fax 81-3-3219-5710, or send an e-mail to imd_sci@
group.shimadzu.co.jp
Al-Tajheezat
Al-Tibbiya
8
Not compliant.
Software revision-up is available to be compliant. Contact manufacturer.
14
Spectro-
photometer
UV-160A
Shimadzu Y2K Site
Contact Shimadzu
If you would like additional information on Shimadzu's Year 2000 compliance, fax 81-3-3219-5710, or send an e-mail to imd_sci@
group.shimadzu.co.jp
Al-Tajheezat
Al-Tibbiya
3
Conditionally compliant. Uses 2-digit year. Requires ROM version 9 for y2k compliance. Product discontinued. Year "00" cannot be set, but instrument will "roll" to "00." Requires ROM update (206-53617-91 for ver9, or 206-53166-91 for earlier versions.
Recommendations: Allow year to roll to 00. Best to keep battery in operating condition. If unit loses power, set unit to 12/31/99, allow to roll to "00" and then reset month and day.
15
Ultrasound System
Concept 2000
(Classified by Company as a Historical Product)
Dynamic Imaging Home Page
Dynamic Imaging Y2K Product List


For further information on Year 2000 product compliance contact :

 
 
 
 
 
 
 
 
 
 
 

Andrew Johnston, Quality Systems Manager
andrew.johnston@
dynamicimaging.co.uk

Al-Madar
4
Does not recognise 2000 as being a leap year.
The calendar date can be set by the user on 29/02/00 and will show the correct dates thereafter.For systems less than 7 years old the software can be upgraded at a cost of £5000 per software version.

For systems more than 7 years old contact Dynamic Imaging for a quotation.
Discounted purchase of a new system can be arranged. Contact Dynamin Imaging for a quotation.
16
Ultrasound System
Concept 1000
(Classified by Company as a Historical Product)
Dynamic Imaging Home Page
Dynamic Imaging Y2K Product List


For further information on Year 2000 product compliance contact :

 
 
 
 
 
 
 
 
 
 
 

Andrew Johnston, Quality Systems Manager
andrew.johnston@
dynamicimaging.co.uk

Al-Madar
2
Does not recognise 2000 as being a leap year.
The calendar date can be set by the user on 29/02/00 and will show the correct dates thereafter. For systems less than 7 years old the software can be upgraded at a cost of £5000 per software version.

For systems more than 7 years old contact Dynamic Imaging for a quotation.
Discounted purchase of a new system can be arranged. Contact Dynamic Imaging for a quotation.
17
Ultrasound System
Concept 500
(Classified by Company as a Historical Product)
Dynamic Imaging Home Page
Dynamic Imaging Y2K Product List


For further information on Year 2000 product compliance contact:

 
 
 
 
 
 
 
 
 
 
 

Andrew Johnston, Quality Systems Manager
andrew.johnston@
dynamicimaging.co.uk

Al-Madar
9
Does not recognise 2000 as being a leap year.
The calendar date can be set by the user on 29/02/00 and will show the correct dates thereafter.For systems less than 7 years old the software can be upgraded at a cost of £5000 per software version.

For systems more than 7 years old contact Dynamic Imaging for a quotation.
Discounted purchase of a new system can be arranged. Contact Dynamic Imaging for a quotation.
18
Adult Ventilator
7200a 
Nellcor Puritan Bennet Incorporation Home Page
Nellcor Puritan Bennet Y2K Site (including a search facility)
Derhally
31 for 7200 series
Not Year 2000 compliant. These versions of the 7200 Series Ventilator will not display, print or output accurate date information beyond December 31, 1999, even if an effort is made to manually enter the year 2000 into the ventilator. If the year 2000 is manually entered into the ventilator, the ventilator keyboard will display "00", but external devices will display, print and output the year as 1900. In addition to this example, other date inconsistencies will also occur. This does not have any adverse impact on safety and effectiveness of the product or on internal calculations and functions. There is no impact on breath delivery.
Upgrade Information: If accurate date display, printout or output is desired beyond December 31, 1999, purchase of an upgrade is required. These upgrades will produce accurate dates for the year 2000 through the year 2079. Please contact your Nellcor Puritan Bennett representative for further information on upgrade options.
19
Computed Tomography Scan
1200 SX
Picker Y2K Site
Al-Riyad
2
Dates may be displayed in Year 2000 incompatible manner. This is a legacy product and is no longer being manufactured.
No Year 2000 upgrade is planned; however, service support will continue.
20
MRI
(Repeat)
Presige 2T
Elscint
(The company has been acquired by Picker)
Philadelphia
1
1. Impossible to enter patient date of birth.
2. Garbage patient age.
3. Improper dates display on monitor of minor functions.
No Charge for Solution Package.
Installation at no charge for Full Service Contract Customers.
Solution in V3.2

Important note: installation of the solution prior to 1 January 2000 cannot be guaranteed for those customers who contact Elscint after 1 July 1999.

21
Electrolyte analyzer
AVL  982
AVL Company Web Site

AVL Company Y2K Site

Ammon
3
Not OK
Obsolete, will not be tested for Y2K compliance.
These products  will no longer be supported by AVL as of Jan 1, 2000.
Contact your local AVL representative to identify a suitable replacement.
22
Ultrasound System
EUB-315
Hitachi Medical Systems Home Page
Hitachi Medical Corporation of America, Hitachi Medical Systems Y2K Site
Contact Hitachi Medical Systems
Hitachi Medical Corporation of America
Hitachi Medical Systems
660 White Plains Road 
Tarrytown, NY 10591-5107 
Phone: (914) 524-9711 or (800) 852-2080 
Fax: (914) 524-9716 
Email: info@
hitachiultrasound.com
 
Not provided by Ministry.
Product is obsolete or beyond a reasonable useful service life and no upgrade will be provided. When the current DATE is in the year 2000 (00) and the LMP is in the year 1999 (99), the values for GA FROM LMP and EDD FROM LMP will not be displayed in the OB measurement display and the OB measurement report display.
No formal corrective measure is available, however, in order to alert the operator of this potential condition, HMCA will issue a caution label to be supplied at the customer’s request to HMCA. TO RECEIVE A CAUTION LABEL:
Please click here to use the online request form.
23
Cell Counter
F-800
Sysmex Home Page
 Sysmex Y2K Site
For more detailed information,
please call the Technical Support  Center at:
1-888-879-7639 or e-mail
 
7
It will need intervention such as a software modification, a one time operator intervention, etc. Special Instructions to follow.
Contact manufacturer.
24
Ultrasonic Scanner
Sonoace 600?
Medison Pty Ltd
Shuqair
3
S/W Version SA600 V1.40
Test Results:
Y2000 Power on = Pass,
Y2000 Power off = Pass,
Y2001 Power on = Pass,
Y2000 Power off = Pass,
30/10/2000 Monday = Not Applicable,
Leap Year OK? = Fail,
9/9/99 OK? = Pass.
Leap year OK with power off but not power on.
Leave off overnight from 28/2 to 29/2/2000.
25
MicroDiffTM (MDTM) Analyzer 
8
Beckman Coulter Home Page
Beckman Coulter Y2K Site
 
2
Software Version: 1.1
Product is Year 2000 compliant for the intended use and primary functionality: minor functionality subject to operational notes. Operational Notes: Operational Notes P/N 4276750 will be available January 1999. Hardware workaround required for AMI/DTX SCAT386 and AMI/ARROW HAUPPAUGE 386(computer off, date will reset to 1980). Inferencing Rule: Two digit years with a value equal to or greater than 80 imply 19XX; those with a value less than 38 imply 20XX. Years 38 thru 79 not recognized.
The listed software versions and higher (if applicable) have been evaluated for Year 2000 compliance. Lower software versions will not be evaluated for Year 2000 compliance and must be upgraded to the listed or higher software versions to ensure Year 2000 status as reported. Customers should contact their Customer Service Representatives for information about upgrades from lower software versions. Customer Year 2000 compliance testing and/or verification is not recommended or supported, and may cause a disruption in the operation of the product that is not related to Year 2000 compliance. Potential disruptions may include loss of calibration, quality control and/or meter information and expiration of on-board reagents. Customers are responsible for consumables and service fees resulting from customer compliance testing.
26
Pulmonary Function Lab
2450
Sensor Medics Y2K Site
 
1
This system requires a SensorMedics software upgrade to meet Y2K compliancy standards. Also, some computers sold with this system may not be Y2K compliant (or compatible with the software upgrade). See the following note about computer compliance verification. Note: Some computers sold with this system may not be Y2K compliant. The customer will need to verify whether or not their computer is Y2K compliant. See the following note for our link to computer compliance verification procedures. Note: Since it is possible for our customers to supply their own computers (not purchased from SensorMedics), it becomes the customer's responsibility in these instances to verify that their computer is Y2K compliant. Click here for instructions on verification procedures. This instrument is not currently marketed.
 
27
Central Monitor
C814
Lohmeier Medizin Elektronik GmbH + Co.
Fritz-Berne-Strase 40 - 81241 Munchen
Tel: ++49 - (0)89 - 83 94 09 - 0
Fax: ++49 - (0)89 - 83 94 09 - 43
e-mail: Lohmeier@
Lohmeier.com
Al-Riyad
1
"This product will possibly have Year 2000 incompatibilities. System will possibly display and print a wrong week-day from 01.01.2001 on."
An e-mail was sent to the manufacturer on 23/2/1999. Response on 24/2/1999, requesting address and fax to send compliance certificates. E-mail sent on 24/2/1999 giving required information. Manufacturer replied on 24/2/1999 by fax. Lohmeier will provide corrected software versions. For availability and update prices, contact the manufacturer.
28
ECG
Monitor
M608
Lohmeier Medizin Elektronik GmbH + Co.
Fritz-Berne-Strase 40 - 81241 Munchen
Tel: ++49 - (0)89 - 83 94 09 - 0
Fax: ++49 - (0)89 - 83 94 09 - 43
e-mail: Lohmeier@
Lohmeier.com
A-Riyad
16
The manufacturer sent a fax detailing Y2K compliance of his products. This particular product "will possibly have Year 2000 incompatibilities. The system will possibly display a wrong week-day from 01.01.2001 on."
An e-mail was sent to the manufacturer on 23/2/1999. Response on 24/2/1999, requesting address and fax to send compliance certificates. E-mail sent on 24/2/1999 giving required information. Manufacturer replied on 24/2/1999 by fax. The manufacturer will provide corrected software versions. For availability and update prices, contact Lohmeier.
29
Ion Chromatograph
LC-10
Shimadzu Y2K Site
Contact Shimadzu
If you would like additional information on Shimadzu's Year 2000 compliance, fax 81-3-3219-5710, or send an e-mail to imd_sci@
group.shimadzu.co.jp
Al-Tajheezat
Al-Tibbiya
1
This product will NOT process the year 2000 with regards to date formating, or leap year recognition. However, it will remain fully functional before, during and after the Year 2000 date change. 
Print Certificate
Requires version 1.40 or later.
30
Electro-
phoresis System
CliniScan 2

Catalogue No. 1260/
1261

Helena Laboratories Y2K Site

Rita Ellerbrook
800-231-5663

 
1
There are two parts to the question of Year 2000 Compliance, software and hardware. The software is written by Helena Laboratories. All software upgrades are available from us at no cost. The latest version of CliniScan 2 software is Rev. H. Older versions of the software have not been verified as being able to handle Year 2000, and an upgrade to Rev. H is recommended.

The CliniScan 2 software has been found to be compliant for the year 2000 with one exception. Stored data can not be retrieved by date in one pass across the millennium. You must retrieve data from before 01/01/2000 in one pass, and data after 12/31/1999 in a second pass. Data can be retrieved across the millennium by patient ID or other parameter, with no regard for date as always. Instructions for data retrieval by date over a range including 1/1/2000 are found at the end of this letter. A summary of the activities that this software performs correctly is listed below.

  • The instrument functions in exactly the same manner before and after 1/1/2000. 
  • It recognizes the year 2000 as a leap year. 
  • It correctly handles all 9s in the data field (9/9/1999 was evaluated). 
  • It correctly handles an 8-digit date in the year 2000 (10/10/2000 was tested). 
  • The date may roll over to a prior year and may then be reset to 2000. 
  • After the year 2000, the date rolls over correctly to the year 2001. 
  • The date format for the 21st century must contain 4 digits (i.e., 2000, 2 digits assume the 20th century). 
The second part of Year 2000 Compliance, hardware, regards the nature of the BIOS chip on the computer board. This chip will determine if the date will automatically change from 12/31/1999 to 01/01/2000 and from 02/28/2000 to 02/29/2000. You will need to check the date on January 1, 2000 and again on February 29, 2000 and, perhaps, manually reset the date. Once this is done, the software will perform as described.

NOTE: Because of the possibility of the date improperly rolling over, do not create data at the time of the date change.

Setting the Date:
1. Press F6 to enter Utilities.
2. Press 2 (Date).
3. Press 1 or 2 for date format.
4. Enter the date as 01-01-2000 and press Enter. You must enter the year as four digits for the 21st century.
5. Press F2 to return to the Test Select menu.

Retrieving Data by Date Range:
1. On the Test Select menu, press A to N for the Test List.
2. Press F5 (Files).
3. Press R, Enter.
4. At date scanned, type the date, such as 01-01-98...* to retrieve data from 1/1/98 through 12/31/99, even if there are data present at later dates. 
5. Press F5 to retrieve the data.
6. To retrieve data after 12/31/1999, enter a date range such as 01-01-00...05-01-00 to retrieve data from 1/1/2000 to 5/1/2000.

7. Press F5 and the data will be retrieved.
 

 
31
 
As stand alone system
Technicon DAX

Technicon DAX (UIW)

Bayer Corporation, Business Group Diagnostics Y2K Site
Richard Kaplan
Phone: 914-631-8000
Techno-
Medix
1
As stand alone system: Customer Bulletin to be made available with instructions by 4/1/1999.
S/W Version 4.03, 4.10
Date Format: mm/dd/yy

Technicon DAX (UIW): Compliant software projected to be released by 4/1/1999.
S/W Version 5.11
Date Format: mm/dd/yy

Contact local agent or manufacturer for solution.
32
Ultrasound
Performa
Dornier 
(Acoustic Imaging)
 
8
Testing/Results

The testing methodology encompassed full system functionality testing both during and after the year 2000 transition. The testing investigated several issues; the rollover and retention of the time and date from 12/31/99 to 1/1/00, the proper identification and use of leap years subsequent to 1/1/00, i.e. 2000, 2004 etc., the real time functionality of the unit including display information and transducer operation, the data archiving functionality of the system, the two dimensional and velocity measurement functionality of the system, and the calculation package functionality of the system

Completion of the aforementioned testing has indicated that subject systems will roll over to January 1, 2000 but will not retain this information after the system is powered off. The testing did indicate however, that once the date and time had been reset to January 1, 2000 or later, then the system would perform as expected including the functionality as indicated above.
 

 

Corrective Action
Each customer must install the enclosed software upgrade version 2.18 to assure the latest and tested version of the software is installed. To install the software, please follow these instructions:
Turn the system off.

Insert the version 2.18 software diskette into the floppy disk drive on the front of the system.

Turn the system on.

The system will load the files from the floppy into the hard drive. This will take approximately 5 minutes. Once the files are loaded, the system will beep.

 
Remove the diskette.

Turn the system off and on again to resume normal function.

Store the version 2.18 software diskette in a safe place.

After installing the new version software, the customer must use the system settings utilities to reset the time and date of their system upon its initial use anytime AFTER 12:00A.M. January 1, 2000 to accurately transition from 1999 to the year 2000 and thereafter. The following steps from the "System Customization" portion of the operators manual explains the proper sequence for setting the time and date.
From the Top Menu of the Soft Key Panel choose the button labeled "OTHER CHOICES".
 
Press the Soft Key button labeled "CH TIME & DATE".
Activation of the Change Time and Date Panel will provide an on screen prompting to guide you through changing the time and date. To enter press Select Key.

Note: It is important that the customer reset the system to the current time and date upon initial use AFTER 12:00 A.M., January 1, 2000.

Conclusion

After performing the aforementioned corrective action, regardless of when any data was taken for system storage (before, during or after the year 2000), the unit will function properly in all imaging, measurement, and calculation conditions.

 33
Immunoassay Instrument
TDX Version 12?
If it is Version 15, then comments herein are valid.
 Abbott Laboratories Y2K Site
 
 2
 Year 2000 compliant with intervention such as software or hardware upgrade or special instructions, etc.
Software upgrade Rev. 16.0 shipping. Contact local Abbott representative.
 34
 Defibrillator Monitor
Defigard
3002 IH 
Bruker Y2K Site

Bruker Products'
Y2K Status

Al-Khadamat Al-Tibbiya Al-Musanida?
20 
The manufacturer sent a certificate to the effect that this system will be partially affected by the transition to the Year 2000.
The solution, according to the manufacturer, is to clear the memory when entering the new millennium.
 35
 Ambulatory pH, Gastro-Intestinal Motility and Urodynamic Systems
 NYMPHNYMPH
 Albyn Medical Ltd.
Bridgend Road Industrial Estate
Dingwall
Ross-shire IVV15 9QF
Scotland
Tel.: +44 1349 862388
Fax: +44 1349 864146
e-mail: valeriel@albyn.com
 
 1
 The NYMPH product range stores the date and time in its memory. The date and time can be set either from the keypad or by downloading from the host computer. When setting the date and time from the keypad, the year is displayed in two digit format only and only years from 80 to 99 are recognised; however, in our experience few users if any set the date from the keypad, most rely on the automatic download from the PC and realistically these data loggers will not present a problem. The analysis software works in the same way as the PHOENIX family and the host computer can be checked in the same way. If you find that there are problems with Year 2000 compliance with our products, contact manufacturer, who will take the matter up and discuss it with the computer manufacturers concerned.
 

 
36
Ultrasound System
AU530
Esaote Home Page

Ultrasound Site

Contact the Company

Jordan Medicare Corporation
Tel.: 5664008
Fax: 5665093
5
Web site does not include y2k information. The manufacturer sent a letter detailing the status of equipment. The following comments apply to this particular model:
1. The date after 1 January 2000 is not correct.
2. The birthdate is not correct crossing the millennium.
3. The LMP date is not correct crossing the millennium.
Contact the manufacturer directly. Company contacted on Wednesday 24/2/1999. 
1. The operator is able to correct the date permanently.
2. The operator has to manually compute the birthdate between 1 January 2000 and 30 September 2000.
3. The operator has to manually compute the LMP date between 1 January 2000 and 30 September 2000.
 37
Anaesthesia Machine
NARKOMED 2B
North American Drager Y2K Site
Y2K Specific Contact Information
3135 Quarry Road
Telford, PA 18969
USA
Tel.: (215) 721 5400
Fax: (215) 723 5934
e-mail: brennanj@nad.com
   
Software version displayed at power on 1.04
Test results 1. (stand alone) pass
2. (data in) N/A
3. (data out) pass
Note: A third party device may request a time and date message which contains a two digit year. N.A.D., Inc. cannot guarantee compliance with other manufacturer's products that are attached to or used with their anaesthesia machines. The Operating Room Data Manager (ORDM) which is commonly attached to a Narkomed Anesthesia machine and utilized with off-line software is not Year 2000 compliant. The ORDM will represent the first two digits of the patient's birth date as '19'. If your software is earlier than that listed in Table 1 we recommend an upgrade of the software and hardware, if needed, to achieve a level of compliance with our published testing protocol. In the U.S. contact DrägerService at (800) 543-5047. Narkomed Anesthesia machines do not utilize date or time information in their life support or drug delivery functions.
 
               
               
               